MATERIALS COMPATIBILITY
- No adverse impact on metallic or non-metallic components of flexible endoscopes
- Non-damaging to electro-plating
- No adverse impact on dental products tested including amalgam
REPRODIS CE HLD4i will not harm sensitive instruments including flexible fibre optic instruments.
APPLICATIONS
To prepare 1 litre of REPRODIS HLD4i ready-to-use solution
and add 250ml of concentrate, for 15 litres use 750ml of concentrate and make up to level with water.
- Prepare 950ml of water
- Add 50ml of REPRODIS CE HLD4i concentrate into 950ml of prepared water
- Stir gently to ensure mixing
- Larger quantities can be prepared using the same ratios, e.g. for 5 litres prepare 4750ml of water
Rigid endoscopes and other instruments/devices
- Immerse in proteolytic enzyme cleaner and wash clear of all debris
- Rinse thoroughly and dry
- Totally immerse in REPRODIS CE HLD4i Instrument Disinfectant for recommended time
- Remove and rinse in filtered or sterile water
- Dry and store in approved conditions
Flexible endoscope cleaning and disinfection with disposable sponges or swabs and clean distal end with appropriate method
- Immerse disassembled endoscope in warm water andproteolytic enzyme cleaner. Wash the outside
- Remove valves, clean and disinfect in REPRODIS CE HLD4i Instrument Disinfectant for recommended contact time
- Brush clean the biopsy-suction channel
- Rinse all parts thoroughly and dry
- Totally immerse the scope in REPRODIS CE HLD4i Instrument Disinfectant for
- Rinse thoroughly with filtered or sterile water, as required
- Dry and store in approved conditions
- If the instrument is to be re-used immediately , purge all rinse water from the channels, dry